Comparison of polymer-tipped and metal-tipped irrigation-aspiration instruments in irrigation and aspiration of anterior chamber Chika Kurosawa 1 , Junya Kizawa 1 , Makoto Komatsu 2 , Daijiro Kurosaka 1 1Dept of Ophthalmol, Iwate Med Univ Sch of Med 2Dept of Physics, Cent for Liberal Arts and Sci, Iwate Med Univ pp.1421-1426
Published Date 2014/10/15
DOI https://doi.org/10.11477/mf.1410200009
  • Abstract
  • Look Inside
  • Reference

Abstract. Purpose:To evaluate whether the polymer-tipped irrigation/aspiration instrument(polymer tip)has better aspiration efficiency and stability of the anterior chamber than metal one(metal tip). Method:Using three types of polymer tips and one type of metal tip, we measured vacuum pressure of aspiration without occlusion, initial aspiration volume(the aspiration volume during 30 seconds after starting)and the irrigated volume. Results:Polymer tips had lower vacuum pressure of aspiration, greater initial aspiration and irrigated volume than metal tip. Conclusion:These findings suggest that polymer tips may have better aspiration efficiency and stability of the anterior chamber than metal tip.


Copyright © 2014, Igaku-Shoin Ltd. All rights reserved.