Coma-like aberration increases after YAG laser application to the center of intraocular lens Maria Tatsugawa 1 , Seiji Tanimoto 2 , Shigeki Yoshitani 1 , Takashi Noma 1 , Yoshiaki Kiuchi 2 1Dept of Ophthalmol,Saiseikai Kure Hosp 2Dept of Ophthalmol and Vis Sci,Grad Sch of Biomed Sci,Hiroshima Univ pp.937-939
Published Date 2009/6/15
DOI https://doi.org/10.11477/mf.1410102754
  • Abstract
  • Look Inside
  • Reference

Abstract. Purpose:To report changes in higher-order aberration following disruption of center of intraocular lens(IOP)by YAG laser. Methods:This experiment was made on 3 types of IOP totaling 14 pieces. A single shot of Q-switched YAG laser was applied to the posterior center of IOP with the power output of 1 mJ. IOL was then placed in a model eye and its aberrations were measured using an OPD-ScanⅡ. Results:IOL showed a pit only and no crack. There was no change in higher-order or spherical-like aberrations in any of 14 IOLs. There was a significant change in coma-like aberration(p=0.04). The amount of change was independent on the type of IOL. Conclusion:Coma-like aberration increases after formation of pit in the posterior surface of IOL by application of YAG laser.


Copyright © 2009, Igaku-Shoin Ltd. All rights reserved.