Flow rate and leakage of irrigating fluid during phacoemulsification with adjunct use of ultrasleeve Nae Kitamura 1 , Hiroko Bissen-Miyajima 1 , Kunihiko Nakamura 2 1Dept of Ophthalmol, Tokyo Dent Coll Suidobashi Hosp 2Tanashi Nakamura Eye Clin pp.1025-1027
Published Date 2007/6/15
DOI https://doi.org/10.11477/mf.1410101802
  • Abstract
  • Look Inside
  • Reference

Abstract. Purpose:To report dynamics of irrigating fluid using ultrasleeve during phacoemulsification-aspiration(PEA). Cases:This study was made on 34 eyes that received PEA with ultrasleeve. Findings were compared with our previous report on PEA using microsleeve. Results:Flow rate of irrigating fluid was significantly less for ultrasleeve than microsleeve. There were no signifcant differences between two groups regarding time of activation of ultrasound, duration of aspiration, or leakage of irrigating fluid. Conclusion:PEA with ultrasleeve results in lesser flow of irrigatinig fluid and smaller corneal incision than PEA with microsleeve.