Evaluation of CoaguChek XS for Japanese Patients: A new procedure of PT-INR measurement Seiichi Taniai 1 , Kyoko Hoshida 1 , Satoru Yusu 1 , Taku Watanabe 2 , Hideaki Yoshino 1 1Division of Cardiology, Second Department of Internal Medicine, Kyorin University School of Medicine 2Department of Clinical Pathology, Kyorin University School of Medicine Keyword: ワルファリン , コアグチェック , PT-INR , warfarin , CoaguChek pp.429-433
Published Date 2010/4/15
DOI https://doi.org/10.11477/mf.1404101466
  • Abstract
  • Look Inside
  • Reference

 Purpose: A new point-of-care(POC) monitoring system of oral anticoagulation, the CoaguChek XS, has recently been developed to measure the international normalized ratio(INR) and may offer significant advantages. The purpose of our study is to compare values obtained by the CoaguChek XS system(COGC-INR) with the standard venous INR for anticoagulant monitoring, prothrombin time with reference thromboplastin(PT-INR).

 Methods and Results: Paired PT-INR and COGC-INR were measured in consecutive 115 samples from 109 patients. Correlation between COGC-INR and PT-INR was r=0.97. The difference in the value of INR deviating from normal distribution of the Bland-Altman plot was 0.3, and values greater than 0.3 were observed in 7 samples. The chi-square test was performed in these 7 samples to investigate the influence of drugs, but did not show any significant influence of drugs on the differences between COGC-INR and PT-INR.

 Conclusions: The accuracy of the CoaguChek XS POC monitoring system when compared with the standard reference INR was satisfactory. This result will improve clinical decision-making regarding anticoagulation therapy.


Copyright © 2010, Igaku-Shoin Ltd. All rights reserved.