40歳,男性.3年前に経皮的腎砕石術を受け,術後に左下腎杯に親水性ポリマーコーティングガイドワイヤー(RADIFOCUS®)の断端8mmの遺残が確認された.経過観察されていたが自然排出なく,異物を核として結石が形成されたため,腎異物結石に対し体外衝撃波結石破砕術を施行した.治療1ヵ月後結石および異物の完全排出を確認した.
A 40-years-old male patient consulted our hospital complaining of the left lumbago. He underwent left percutaneous nephrolithotomy at another hospital 3 years before. X-ray examinations showed a soybean-sized foreign body stone originated from the hydrophilic polymer coating of guidewire (RADIFOCUS®) used and left in the lower area of the left kidney at the previus operation. We treated a patient with extracorporeal shock wave lithotripsy (ESWL). Stone and foreign body were disintegrated, and 1 month later after ESWL plain X-ray film showed no stone and foreign body.