Effect of Magnetic Fields from Home-use Magnetic Induction Therapy Apparatuses on Adjustable Cerebrospinal Fluid Shunt Valves Koji NAKASHIMA 1 , Atsunori OISHI 1 , Hiroshi ITOKAWA 2 , Michio FUJIMOTO 3 1Department of Neurosurgery,Higashi-Totsuka Memorial Hospital 2Department of Neurosurgery,Kasai Cardiology and Neurosurgery Hospital 3Department of Neurosurgery,Showa University Fujigaoka Hospital Keyword: hydrocephalus , magnetic field , shunt valve , safety pp.725-729
Published Date 2010/8/10
DOI https://doi.org/10.11477/mf.1436101226

閲覧方法のご案内

この論文を閲覧するためにはログインが必要になります。
パスワードを忘れた場合 新しくユーザー登録する 施設共通IDをご利用の方はこちらから

PPV購入案内

1,320円 (1,200円+税10%) こちらは、725ページから729ページの文献です。
PPV(ペイ・パー・ビュー)とは、論文(記事)単位で購入・閲覧ができるサービスです。
購入には医書.jp本会員登録が必要です。



会員登録完了後に改めて上記「購入サイトへ」を選択してください。
または「購入サイトへ」選択後に「購入ログイン」、「新規会員登録」の順に進んでください。

会員登録について詳しくはをご確認ください。
  • Abstract
  • Look Inside
  • Reference

 Objective: Cerebrospinal fluid (CSF) shunts are frequently used to treat hydrocephalus. The use of a programmable valve allows the operator to easily change the opening pressure. In Japan, many people use magnetic induction therapy apparatuses in their homes. However, exposing patients with adjustable CSF shunt valves to the permanent magnets included in these apparatuses may alter the shunt valve's programmed settings or permanently damage the device. Therefore, the goal of this study was to determine the health risk associated with magnetic induction therapy for patients using programmable CSF shunt valves.

 Methods: Five models of shunt valves from five different manufacturers, the Miethke proGAV (proGAV), the Codman Hakim programmable valve (CHPV), Sophysa Sophy model SM8 (Sophy valve), Sophysa Polaris model SPV (Polaris valve), and StrataⅡ valve (Strata valve) were evaluated in this study. Magnetic field interactions were determined for the programmable valves by using magnetic stones with various magnetic flux densities. The maximum distance between the valve and the magnetic stone affecting the valve pressure setting was measured by X-ray.

 Results: The proGAV and Polaris valve were immune to unintentional reprogramming by the magnetic stones. The CHPV, Sophy valve and Strata valve, however, randomly changed settings by magnetic stones.

 Conclusions: Whereas the CHPV, Sophy valve and Strata valve were promptly reset by exposure to a magnetic stone with a similar strength to that used in magnetic induction therapy, proGAV and Polaris valve were resistant to inadvertent reprogramming when exposed to magnets up to 190mT.