Neck Fracture of an Anatomic Medullary Locking Plus Stem. Case Report Tomonori MISAKI 1 , Yasuhiro UEDA 1 , Masayuki HAYASHI 1 , Motoi ISHIGURO 1 , Hiroshige HAMADA 1 , Hiroyuki NAKANISHI 1 , Yasuhisa YOSHIDA 1 , Atsushi MURATA 2 1Department of Orthopaedic Surgery, Fukui Prefectural Hospital 2Department of Orthopaedic Surgery, Fukui Prefectural Rehabilitation Center for Disabled Children Keyword: 人工股関節全置換術 , THA , 頚部折損 , neck fracture , AML plusステム , AML plus stem pp.1241-1245
Published Date 2013/12/25
DOI https://doi.org/10.11477/mf.1408102915
  • Abstract
  • Look Inside
  • Reference

 Primary total hip arthroplasty (THA) with anatomic medullary locking (AML) stem have shown long-term success. We report a case of neck fracture of an AML plus stem 15 years after primary THA. A 79-year-old man suddenly experienced right hip pain while trying to stand up in a garden. A radiography showed neck fracture at the trunnion of the stem, and a CT scan showed a retroverted cup and massive osteolysis in the acetabulum. We performed revision THA using extended trochanteric osteotomy, and he is doing well 1 year after surgery. Sixteen neck fracture cases of the AML plus stem have been reported in Japan, and these cases occurred most frequently when 28-mm head and +6-mm neck was used in combination. We must pay particular attention when an AML plus stem with this combination is used.